The Microstructure And The Electrochemical Behavior Of Cobalt Chromium Molybdenum Alloys From Retrieved Hip Implants, Christopher P. Emerson May 2015

The Microstructure And The Electrochemical Behavior Of Cobalt Chromium Molybdenum Alloys From Retrieved Hip Implants, Christopher P. Emerson

FIU Electronic Theses and Dissertations

Because of their excellent mechanical, tribological, and electrochemical properties, Cobalt Chromium Molybdenum alloys have been used as the material for both the stem and head of modular hip implants. Corrosion is one mechanism by which metal debris, from these implants, is generated, which can lead to adverse events that requires revision surgery. Manufacturing process such as wrought, as-cast, and powder metallurgy influences the microstructure, material properties, and performance of these implants

The current research focuses on analyzing the microstructure of CoCrMo alloys from retrieved hip implants with optical and scanning electron microscopy. Additionally, energy disperse spectroscopy was utilized to determine …


In-Vivo Corrosion And Fretting Of Modular Ti-6al-4v/Co-Cr-Mo Hip Prostheses: The Influence Of Microstructure And Design Parameters, Jose Luis Gonzalez Jr Apr 2015

In-Vivo Corrosion And Fretting Of Modular Ti-6al-4v/Co-Cr-Mo Hip Prostheses: The Influence Of Microstructure And Design Parameters, Jose Luis Gonzalez Jr

FIU Electronic Theses and Dissertations

The purpose of this study was to evaluate the incidence of corrosion and fretting in 48 retrieved titanium-6aluminum-4vanadium and/or cobalt-chromium-molybdenum modular total hip prosthesis with respect to alloy material microstructure and design parameters. The results revealed vastly different performance results for the wide array of microstructures examined. Severe corrosion/fretting was seen in 100% of as-cast, 24% of low carbon wrought, 9% of high carbon wrought and 5% of solution heat treated cobalt-chrome. Severe corrosion/fretting was observed in 60% of Ti-6Al-4V components. Design features which allow for fluid entry and stagnation, amplification of contact pressure and/or increased micromotion were also shown …


An Assessment Of Novel Biodegradable Magnesium Alloys For Endovascular Biomaterial Applications, Dharam Persaud-Sharma Jun 2013

An Assessment Of Novel Biodegradable Magnesium Alloys For Endovascular Biomaterial Applications, Dharam Persaud-Sharma

FIU Electronic Theses and Dissertations

Magnesium alloys have been widely explored as potential biomaterials, but several limitations to using these materials have prevented their widespread use, such as uncontrollable degradation kinetics which alter their mechanical properties. In an attempt to further the applicability of magnesium and its alloys for biomedical purposes, two novel magnesium alloys Mg-Zn-Cu and Mg-Zn-Se were developed with the expectation of improving upon the unfavorable qualities shown by similar magnesium based materials that have previously been explored. The overall performance of these novel magnesium alloys has been assessesed in three distinct phases of research: 1) analysing the mechanical properties of the as-cast …


Electrochemical Investigation Of Chromium Nanocarbide Coated Ti-6al-4v And Co-Cr-Mo Alloy Substrates, Viswanathan Swaminathan, Haitong Zeng, Daniel Lawrynowicz, Zongtao Zhang, Jeremy L. Gilbert Jan 2011

Electrochemical Investigation Of Chromium Nanocarbide Coated Ti-6al-4v And Co-Cr-Mo Alloy Substrates, Viswanathan Swaminathan, Haitong Zeng, Daniel Lawrynowicz, Zongtao Zhang, Jeremy L. Gilbert

Biomedical and Chemical Engineering - All Scholarship

This study investigated the electrochemical behavior of chromium nano-carbide cermet coating applied on Ti–6Al–4V and Co–Cr–Mo alloys for potential application as wear and corrosion resistant bearing surfaces. The cermet coating consisted of a highly heterogeneous combination of carbides embedded in a metal matrix. The main factors studied were the effect of substrate (Ti–6Al–4V vs. Co–Cr–Mo), solution conditions (physiological vs. 1 M H2O2 of pH 2), time of immersion (1 vs. 24 h) and post coating treatments (passivation and gamma sterilization). The coatings were produced with high velocity oxygen fuel (HVOF) thermal spray technique at atmospheric conditions to …


Focal Osteolysis At The Junctions Of A Modular Stainless-Steel Femoral Intramedullary Nail, Darron M. Jones, Lawrence Marsh, James V. Nepola, Joshua J. Jacobs, Anastasia K. Skipor, Robert M. Urban, Jeremy L. Gilbert, Josheph A. Buckwater Jan 2001

Focal Osteolysis At The Junctions Of A Modular Stainless-Steel Femoral Intramedullary Nail, Darron M. Jones, Lawrence Marsh, James V. Nepola, Joshua J. Jacobs, Anastasia K. Skipor, Robert M. Urban, Jeremy L. Gilbert, Josheph A. Buckwater

Biomedical and Chemical Engineering - All Scholarship

Background:

During routine follow-up of patients treated with a three-piece stainless-steel modular femoral nail, osteolysis and periosteal reaction around the modular junctions of some of the nails were noted on radiographs. The purpose of this study was to evaluate the prevalence, etiology, and clinical relevance of these radiographic findings.

Methods:

Forty-four femoral fractures or nonunions in forty-two patients were treated with a modular stainless-steel femoral intramedullary nail. Seventeen nails were excluded, leaving twenty-seven intramedullary nails in twenty-seven patients for this study. All patients had had a femoral diaphyseal fracture; nineteen had had an acute fracture and eight, a nonunion. These …


Current Concepts Review - Corrosion Of Metal Orthopaedic Implants, Joshua J. Jacobs, Jeremy L. Gilbert, Robert M. Urban Jan 1998

Current Concepts Review - Corrosion Of Metal Orthopaedic Implants, Joshua J. Jacobs, Jeremy L. Gilbert, Robert M. Urban

Biomedical and Chemical Engineering - All Scholarship