Electrochemical Enzyme Immunoassay Using Sequential Saturation Technique In A 20-Μl Capillary: Digoxin As A Model Analyte, Noriaki Kaneki, Yan Xu, Anu Kumari, H. Brian Halsall, William R. Heineman, Peter T. Kissinger Mar 1994

Electrochemical Enzyme Immunoassay Using Sequential Saturation Technique In A 20-Μl Capillary: Digoxin As A Model Analyte, Noriaki Kaneki, Yan Xu, Anu Kumari, H. Brian Halsall, William R. Heineman, Peter T. Kissinger

Chemistry Faculty Publications

Capillary enzyme immunoassay with flow-injection analysis for digoxin using the sequential saturation technique has been developed. Glass capillary tubes (10 cm × 0.53 mm i.d.) with immobilized digoxin antibody were used as the immunoassay reactor. The product of enzymatic reaction. 4-aminophenol, was detected amperometrically. The digoxin and the labeled digoxin binding reaction with the immobilized digoxin antibody were completed in 2 and 10 min, respectively. Digoxin was determined in a 20-μl sample with a detection limit of 10 pg ml−1 (200 fg or 260 attomoles) and a 3 orders of magnitude range.