Feasibility Of Using Ring-Mounted Halcyon Linac For Single-Isocenter/Two-Lesion Lung Stereotactic Body Radiation Therapy, Damodar Pokhrel, Aaron Webster, Richard Mallory, Justin Visak, Mark E. Bernard, Ronald C. Mcgarry, Mahesh Kudrimoti Feb 2022

Feasibility Of Using Ring-Mounted Halcyon Linac For Single-Isocenter/Two-Lesion Lung Stereotactic Body Radiation Therapy, Damodar Pokhrel, Aaron Webster, Richard Mallory, Justin Visak, Mark E. Bernard, Ronald C. Mcgarry, Mahesh Kudrimoti

Radiation Medicine Faculty Publications

PURPOSE: To demonstrate the plan quality and delivery efficiency of volumetric-modulated arc therapy (VMAT) with the Halcyon Linac ring delivery system (RDS) in the treatment of single-isocenter/two-lesion lung stereotactic body radiation therapy (SBRT).

MATERIALS/METHODS: Sixteen previously treated non-coplanar VMAT single-isocenter/two-lesion lung SBRT plans delivered with SBRT-dedicated C-arm TrueBeam Linac were selected. Prescribed dose was 50 Gy to each lesion over five fractions with treatment delivery every other day and AcurosXB algorithm as the final dose calculation algorithm. TrueBeam single-isocenter plans were reoptimized for Halcyon Linac with coplanar geometry. Both TrueBeam and Halcyon plans were normalized for identical combined target coverage …