Open Access. Powered by Scholars. Published by Universities.®
Department of Pathology, Anatomy, and Cell Biology Posters
Activated Clotting Time (ACT): Comparison of the Hemochron Signature Elite and the Abbott i-STAT
Articles 1 - 1 of 1
Full-Text Articles in Medical Cell Biology
Activated Clotting Time (Act): Comparison Of The Hemochron Signature Elite And The Abbott I-Stat, Lilah Evans, Susanne Gallo, Joseph J. Leo, Barbara Goldsmith, Phd, Facb
Activated Clotting Time (Act): Comparison Of The Hemochron Signature Elite And The Abbott I-Stat, Lilah Evans, Susanne Gallo, Joseph J. Leo, Barbara Goldsmith, Phd, Facb
Department of Pathology, Anatomy, and Cell Biology Posters
BACKGROUND
ACT is commonly used for heparin anticoagulation monitoring during procedures including cardiopulmonary bypass surgery, coronary angioplasty, and interventional radiology. To prevent thrombosis, moderate to high levels of heparin anticoagulation are required. The Hemochron Signature Elite (HSE, Accriva Diagnostics, formerly ITC, San Diego, CA) was implemented at TJUH as a replacement for the older model Hemochron Response (Accriva/ITC Model HRS.110, San Diego, CA). Operating room (OR) perfusionists reported irreproducible high results using HSE that could not be explained clinically. In consideration of use of i-STAT analyzers (Abbott Point of Care, Princeton, NJ) as an alternative to HSE, we performed a …