Open Access. Powered by Scholars. Published by Universities.®

2010

Anthropomorphic phantom

Articles 1 - 1 of 1

Full-Text Articles in Other Analytical, Diagnostic and Therapeutic Techniques and Equipment

Commissioning An Anthropomorphic Spine And Lung Phantom For Remote Dose Verification Of Institutions Participating In Rtog 0631, Douglas F. Caruthers May 2010

Commissioning An Anthropomorphic Spine And Lung Phantom For Remote Dose Verification Of Institutions Participating In Rtog 0631, Douglas F. Caruthers

Dissertations & Theses (Open Access)

The RPC developed a new phantom to ensure comparable and consistent radiation administration in spinal radiosurgery clinical trials. This study assessed the phantom’s dosimetric and anatomic utility. The ‘spine phantom’ is a water filled thorax with anatomy encountered in spinal radiosurgery: target volume, vertebral column, spinal canal, esophagus, heart, and lungs. The dose to the target volume was measured with axial and sagittal planes of radiochromic film and thermoluminescent dosimeters (TLD). The dose distributions were measured with the radiochromic film calibrated to the absolute dose measured by the TLD. Four irradiations were administered: a four angle box plan, a seven …