Ultrasound Three- Dimensional Velocity Measurements By Feature Tracking, Gregory R. Bashford, Olaf T. Von Ramm May 1996

Ultrasound Three- Dimensional Velocity Measurements By Feature Tracking, Gregory R. Bashford, Olaf T. Von Ramm

Biomedical Imaging and Biosignal Analysis Laboratory

This article describes a new angle-independent method suitable for three-dimensional (3-D) blood flow velocity measurement that tracks features of the ultrasonic speckle produced by a pulse echo system. In this method, a feature is identified and followed over time to detect motion. Other blood flow velocity measurement methods typically estimate velocity using one- (1-D) or two-dimensional (2-D) spatial and time information. Speckle decorrelation due to motion in the elevation dimension may hinder this estimate of the true 3-D blood flow velocity vector. Feature tracking is a 3-D method with the ability to measure the true blood velocity vector rather than …