Evaluation Of The Quantitative Accuracy Of A Commercially-Available Positron Emission Mammography Scanner, Adam Springer Aug 2010

Evaluation Of The Quantitative Accuracy Of A Commercially-Available Positron Emission Mammography Scanner, Adam Springer

Dissertations & Theses (Open Access)

Objective: The PEM Flex Solo II (Naviscan, Inc., San Diego, CA) is currently the only commercially-available positron emission mammography (PEM) scanner. This scanner does not apply corrections for count rate effects, attenuation or scatter during image reconstruction, potentially affecting the quantitative accuracy of images. This work measures the overall quantitative accuracy of the PEM Flex system, and determines the contributions of error due to count rate effects, attenuation and scatter.

Materials and Methods: Gelatin phantoms were designed to simulate breasts of different sizes (4 – 12 cm thick) with varying uniform background activity concentration (0.007 – 0.5 μCi/cc), cysts and …