Fatigue Testing And Computational Analysis Of A Spinal Implant, Timothy Helton Jan 2011

Fatigue Testing And Computational Analysis Of A Spinal Implant, Timothy Helton

Electronic Theses and Dissertations

The following study investigates the fatigue behavior of a spinal implant developed by Synthes, Inc. to provide stability and off load pressure from the discs in the lumbar region. The installation process for the StenoFix design utilizes plastic deformations to customize the device to each individual patient, but also has potential to degrade fatigue behavior. Physical testing of two titanium alloys, Ti-6Al-7Nb and Ti-15Mo, was conducted on hydraulic test frames and compared to computational simulations which were carried out for the following scenarios: cyclic excitation following plastic deformation; variation of material property definition; variation of loading location; and three modifications …