The Influence Of Acoustic Impedance Mismatch On Post-Stenotic Pulsed-Doppler Ultrasound Measurements In A Coronary Artery Model, S. A. Jones, H. Leclerc, George P. Chatzimavroudis, Y. H. Kim, N. A. Scott, Ajit P. Yoganathan Jan 1996

The Influence Of Acoustic Impedance Mismatch On Post-Stenotic Pulsed-Doppler Ultrasound Measurements In A Coronary Artery Model, S. A. Jones, H. Leclerc, George P. Chatzimavroudis, Y. H. Kim, N. A. Scott, Ajit P. Yoganathan

Chemical & Biomedical Engineering Faculty Publications

Acoustic impedance mismatch at the fluid-wall interface was shown to affect the spectra from an intravascular Doppler device in an in vitro model with a diameter typical of human coronary arteries. Measurements were obtained first under Poiseuille flow conditions with impedance mismatches of 0%, 7% and 12%, and then under stenosed conditions for the 0% and 7% mismatch cases. For the zero mismatch case, the Doppler spectra could be readily interpreted in terms of fluid mechanical phenomena. When mismatch was present, the spectra from Poiseuille flow exhibited multiple peaks which could not be directly related to the velocity profile. Also, …