Corrosion Of Metal Hip Arthroplasties And Its Possible Role In Loosening, H. K. Koerten, J. J. A. M. Van Raay, J. J. M. Onderwater, F. P. Bernoski, P. M. Rozing Jan 1997

Corrosion Of Metal Hip Arthroplasties And Its Possible Role In Loosening, H. K. Koerten, J. J. A. M. Van Raay, J. J. M. Onderwater, F. P. Bernoski, P. M. Rozing

Cells and Materials

Tissue removed with human hip arthroplasties during revision surgery of 45 patients was evaluated by transmission electron microscopy and X -ray microanalysis (XRMA). The results show that microscopic and submicroscopic particles are abundantly present in the tissue at the tissue/implant interface. XRMA of individual particles shows that the chemical composition of a portion of the particles was in agreement with that of the retrieved implants. Regularly, particulates with a dissimilar chemical composition were found. Sometimes, these particles could be recognized as filler particles of the cements used. Other particles could partly or completely be composed of the chemical elements that …