Evaluation Of Cirs String Doppler Phantom As A Test Tool For Use In A Doppler Ultrasound Quality Assurance Program, Jacinta Browne, S. Cournane, Andrew Fagan Oct 2012

Evaluation Of Cirs String Doppler Phantom As A Test Tool For Use In A Doppler Ultrasound Quality Assurance Program, Jacinta Browne, S. Cournane, Andrew Fagan

Articles

Ultrasound Doppler systems are routinely used to perform blood flow velocity measurements which assist in the clinical assessment and diagnosis of vascular. Doppler measurements of peak velocities for vascular applications provide an indication of the degree of the stenosis which will ultimately assist in deciding how a patient is managed. It is imperative that Doppler systems are capable of accurately measuring blood flow velocities to ensure correct diagnosis and appropriate patient treatment; therefore such systems should be evaluated regularly as part of a Quality Assurance program. Although a range of Doppler test phantoms have been developed for quality control (QC) …


The Effects Of Fatty Desposits On The Accuracy Of The Fibroscan Liver Transient Elastography Ultrasound System, S. Cournane, Jacinta Browne, Andrew Fagan Jan 2012

The Effects Of Fatty Desposits On The Accuracy Of The Fibroscan Liver Transient Elastography Ultrasound System, S. Cournane, Jacinta Browne, Andrew Fagan

Articles

A new generation of ultrasound transient elastography (TE) systems have emerged which exploit the well-known correlation between the liver’s pathological and mechanical properties through measurements of the Young’s elastic modulus; however, little work has been carried out to examine the effect that fatty deposits may have on the TE measurement accuracy. An investigation was carried out on the effects on the measurement accuracy of a transient elastography ultrasound system, the Fibroscan®, caused by overlaying fat layers of varying thickness on healthy liver-mimicking phantoms, simulating in vivo conditions for obese patients. Furthermore, a steatosis effect similar to that in non-alcoholic fatty …