Open Access. Powered by Scholars. Published by Universities.®

Radiation Medicine Commons

Open Access. Powered by Scholars. Published by Universities.®

Louisiana State University

Margin

Articles 1 - 1 of 1

Full-Text Articles in Radiation Medicine

Analytical Setup Margin For Spinal Sbrt Based On Measured Errors, Audrey Copeland Jun 2020

Analytical Setup Margin For Spinal Sbrt Based On Measured Errors, Audrey Copeland

LSU Master's Theses

Purpose: No consensus currently exists in the radiotherapy community about the correct margin size to use for spinal SBRT. Margins have been proposed to account for various errors individually, but not with all errors combined to result in a single margin value. The purpose of this work was to determine a setup margin for spinal SBRT based on known and measurable errors during radiotherapy to achieve at least 90% coverage of the clinical target volume (CTV) with the prescription dose for at least 90% of patients and not exceed a 30 Gy point dose or 23 Gy to 10% of …