Oxygen Supplementing, Biocompatible Outer Membranes For Enhanced Performance Of Implantable Glucose Sensors, Hardeep Singh May 2010

Oxygen Supplementing, Biocompatible Outer Membranes For Enhanced Performance Of Implantable Glucose Sensors, Hardeep Singh

Honors Scholar Theses

Lack of linearity and sensitivity, oxygen dependence, biofouling and tissue inflammation hinder the development of implantable biosensors for continuous monitoring of glucose. Herein, we report the development of stacked outer membranes based on LBL/PVA hydrogels that improve sensor sensitivity, linearity, oxygen independence and counter biofouling and inflammation. While the inner LBL membrane affords tunable diffusivity, the outer PVA is capable of releasing anti-inflammatory drugs/tissue response modifying agents to counter acute and chronic inflammation, and to induce neo-angiogenesis at the implant site. Sensors were fabricated by immobilizing GOx enzyme on top of 50 μm platinum wires, followed by deposition of stacked …