Monte Carlo Study Of The Energy Response And Depth Dose Water Equivalence Of The Moskin Radiation Dosimeter At Clinical Kilovoltage Photon Energies, C Lian, M Othman, D Cutajar, M Butson, Susanna Guatelli, Anatoly Rosenfeld Nov 2012

Monte Carlo Study Of The Energy Response And Depth Dose Water Equivalence Of The Moskin Radiation Dosimeter At Clinical Kilovoltage Photon Energies, C Lian, M Othman, D Cutajar, M Butson, Susanna Guatelli, Anatoly Rosenfeld

Dean Cutajar

Skin dose is often the quantity of interest for radiological protection, as the skin is the organ that receives maximum dose during kilovoltage X-ray irradiations. The purpose of this study was to simulate the energy response and the depth dose water equivalence of the MOSkin radiation detector (Centre for Medical Radiation Physics (CMRP), University of Wollongong, Australia), a MOSFET-based radiation sensor with a novel packaging design, at clinical kilovoltage photon energies typically used for superficial/orthovoltage therapy and X-ray CT imaging. Monte Carlo simulations by means of the Geant4 toolkit were employed to investigate the energy response of the CMRP MOSkin …


Monte Carlo Study Of Mosfet Packaging, Optimised For Improved Energy Response: Single Mosfet Filtration, M Othman, D Cutajar, Nicholas Hardcastle, S Guatelli, Anatoly Rosenfeld Jan 2012

Monte Carlo Study Of Mosfet Packaging, Optimised For Improved Energy Response: Single Mosfet Filtration, M Othman, D Cutajar, Nicholas Hardcastle, S Guatelli, Anatoly Rosenfeld

Dean Cutajar

Monte Carlo simulations of the energy response of a conventionally packaged single metal-oxide field effect transistors (MOSFET) detector were performed with the goal of improving MOSFET energy dependence for personal accident or military dosimetry. The MOSFET detector packaging was optimised. Two different ‘drop-in’ design packages for a single MOSFET detector were modelled and optimised using the GEANT4 Monte Carlo toolkit. Absorbed photon dose simulations of the MOSFET dosemeter placed in free-air response, corresponding to the absorbed doses at depths of 0.07 mm (Dw(0.07)) and 10 mm (Dw(10)) in a water equivalent phantom of size 30 x 30 x30 cm3 for …