Improvements In Dose Calculation Accuracy For Small Off-Axis Targets In High Dose Per Fraction Tomotherapy, Nicholas Hardcastle, Adam Bayliss, Jeannie Wong, Anatoly B. Rosenfeld, Wolfgang A. Tome Jun 2013

Improvements In Dose Calculation Accuracy For Small Off-Axis Targets In High Dose Per Fraction Tomotherapy, Nicholas Hardcastle, Adam Bayliss, Jeannie Wong, Anatoly B. Rosenfeld, Wolfgang A. Tome

Anatoly Rozenfeld

Purpose: A recent field safety notice from TomoTherapy detailed the underdosing of small, off-axis targets when receiving high doses per fraction. This is due to angular undersampling in the dose calculation gantry angles. This study evaluates a correction method to reduce the underdosing, to be implemented in the current version (v4.1) of the TomoTherapy treatment planning software. Methods: The correction method, termed "Super Sampling" involved the tripling of the number of gantry angles from which the dose is calculated during optimization and dose calculation. Radiochromic film was used to measure the dose to small targets at various off-axis distances receiving …