Tracking Plasma Lactate Concentration In Vivo With A Catheter-Tip L-Lactate Sensor, Brett T. Weinzapfel, Mark D. Ball, Lee R. Waite, Nacer E. Abrouk, Shun P. Lim Mar 1993

Tracking Plasma Lactate Concentration In Vivo With A Catheter-Tip L-Lactate Sensor, Brett T. Weinzapfel, Mark D. Ball, Lee R. Waite, Nacer E. Abrouk, Shun P. Lim

Mathematical Sciences Technical Reports (MSTR)

To circumvent the problems of repeated blood sampling for in vitro analysis, a catheter-tip L-lactate sensor has been developed. The sensor was tested in anesthetized pigs (n=6). The sensor in vivo tracked the lactate concentration non-linearly, seeming to obey Michaelis-Menten kinetics. Calibration time was short, typically 1.5 min per lactate standard. Furthermore, time drift was small, typically -1.3% to -3.3% per hour of in vivo use.