Performance Of A Bgo Pet/Ct With Higher Resolution Pet Detectors, John W. Wilson, Timothy G. Turkington, J. J. Williams, James G. Colsher, D. L. Mcdaniel, C. L. Kim, Steven G. Ross, C. W. Stearns, S. D. Wollenweber Jan 2005

Performance Of A Bgo Pet/Ct With Higher Resolution Pet Detectors, John W. Wilson, Timothy G. Turkington, J. J. Williams, James G. Colsher, D. L. Mcdaniel, C. L. Kim, Steven G. Ross, C. W. Stearns, S. D. Wollenweber

Mining Engineering Faculty Research & Creative Works

A new PET detector block has been designed to replace the standard detector of the Discovery ST PET/CT system. The new detector block is the same size as the original, but consists of an 8/spl times/6 (tangential× axial) matrix of crystals rather than the original 6/spl times/6. The new crystal dimensions are 4.7× 6.3× 30 mm/sup 3/ (tangential× axial× radial). Full PET/CT systems have been built with these detectors (Discovery STE). Most other aspects of the system are identical to the standard Discovery ST, with differences including the low energy threshold for 3D imaging (now 425 keV) and front-end electronics. …


Image Quality Vs. Nec In 2d And 3d Pet, John W. Wilson, Timothy G. Turkington, Josh M. Wilson, James G. Colsher, Steven G. Ross Jan 2005

Image Quality Vs. Nec In 2d And 3d Pet, John W. Wilson, Timothy G. Turkington, Josh M. Wilson, James G. Colsher, Steven G. Ross

Mining Engineering Faculty Research & Creative Works

To investigate the relationship between NEC and image quality to 2D and 3D PET, while simultaneously optimizing 3D low energy threshold (LET), we have performed a series of phantom measurements. The phantom consisted of 46 1 cm fillable hollow spheres on a random grid inside a water-filled oval cylinder, 21 cm tall, 36 cm wide, and 40 cm long. The phantom was imaged on a Discovery ST PET/CT system (GE Healthcare, Milwaukee, WI) in a series of 3 min scans as it decayed from an activity of 7.2 mCi. The scans included LET settings of 375,400, and 425 keV in …