Investigating The Optimum Lower Energy Threshold Of A New Research Pet/Ct Scanner, Osama Mawlawi, John W. Wilson, Tinsu Pan, Timothy G. Turkington Jan 2005

Investigating The Optimum Lower Energy Threshold Of A New Research Pet/Ct Scanner, Osama Mawlawi, John W. Wilson, Tinsu Pan, Timothy G. Turkington

Mining Engineering Faculty Research & Creative Works

An investigation of the optimum 3D lower energy threshold (LET) setting of the Discovery-RX, a new LYSO based GE research PET/CT scanner is conducted. Sensitivity and noise equivalent count rate (NECR) performance of the scanner in 3D mode were evaluated at multiple LET settings: 400, 425, 450, 460, 470 and 480 KeV. The performance evaluations were conducted according to the NEMA NU2-2001 standard. In addition, the NECR was also evaluated for the same LET settings using the Data Spectrum whole body phantom in order to more accurately simulate a true clinical setting. For the sensitivity measurements, the line source was …