Motion And Crosslinked Polyethylene Wear In Reverse Total Shoulder Arthroplasty, Christopher Millward Jul 2022

Motion And Crosslinked Polyethylene Wear In Reverse Total Shoulder Arthroplasty, Christopher Millward

Electronic Thesis and Dissertation Repository

The reverse total shoulder arthroplasty (RTSA) has quickly grown to become the most commonly used shoulder arthroplasty design; however, reports have shown evidence of RTSA failures related to polyethylene wear and damage. Therefore, the present work investigated the wear of crosslinked polyethylene (XLPE) in environments similar to that of an in vivo RTSA. Additionally, a computational model was developed based on a previous study of the shoulder motions obtained from a selection of typical patients with RTSA. This model quantified the amount of glenohumeral motion that an RTSA may be subjected to in vivo and provided an approximate value for …


Influence Of Scapular Notching On Contact Mechanics And Simulator Wear Of Reverse Total Shoulder Arthroplasty Implants, Michael William Griffiths Oct 2017

Influence Of Scapular Notching On Contact Mechanics And Simulator Wear Of Reverse Total Shoulder Arthroplasty Implants, Michael William Griffiths

Electronic Thesis and Dissertation Repository

Scapular notching is a common complication of reverse total shoulder arthroplasty (RTSA) wherein the predominant focus of current literature has been on changes in osseous anatomy. However, the implications on RTSA performance from the damaged humeral cup is largely unknown. Therefore the present work describes the effects of the initiation and propagation of the humeral cup defect resulting from scapular notching through the use of finite element modeling and wear simulation, in order to assess changes in RTSA contact mechanics and tribological properties. A significant decrease in articular contact area and increase in maximum contact stress values was found for …