Open Access. Powered by Scholars. Published by Universities.®

Engineering Science and Materials Commons

Open Access. Powered by Scholars. Published by Universities.®

Power and Energy

2006

Biosensor

Articles 1 - 1 of 1

Full-Text Articles in Engineering Science and Materials

Pre-Oxidative Amperometric Glucose Biosensor Incorporated With Carbon Nanotube And Pbo_(2), Hong-Lan Qi, Cheng-Xiao Zhang Aug 2006

Pre-Oxidative Amperometric Glucose Biosensor Incorporated With Carbon Nanotube And Pbo_(2), Hong-Lan Qi, Cheng-Xiao Zhang

Journal of Electrochemistry

A pre-oxidative amperometric glucose biosensor incorporated with carbon nanotube,ferrocenemonocarboxylic acid and PbO_(2) was designed.The interferences coming from reduced substances,such as ascoribic acid and urine acid were eliminated by PbO_(2).The oxidative current is linear at the potential of +0.40 V(vs.SCE) to the concentration of glucose in the range from 0.50 to 20 mmol·L~(-1),and the detection limit of glucose is 0.1 mol L~(-1).The relative standard deviation is 3.3% for seven successive assays at 5.0 mmol·L~(-1)glucose.This biosensors have been applied to determination of the glucose in serum with satisfying results.