Efficacy And Safety Of The Reciprocal Pulse Defibrillator Current Waveform, M J. Niebauer, Charles F. Babbs, L A. Geddes, J D. Bourland Jan 1984

Efficacy And Safety Of The Reciprocal Pulse Defibrillator Current Waveform, M J. Niebauer, Charles F. Babbs, L A. Geddes, J D. Bourland

Weldon School of Biomedical Engineering Faculty Publications

The efficacy and safety of a new defibrillating current waveform, consisting of a low-tilt 5 ms trapezoidal pulse followed closely by a second identical pulse of opposite polarity, was tested m seven isolated, perfused, working canine hearts suspended in an isoresistive, isosmotic shock bath at 37 oC. The efficacy and safety of the reciprocal pulse was compared with a single 5 ms pulse, a single 10 ms pulse, and a dual (unidirectional) 5 ms pulse waveform. The mean threshold average current densities for the 5 ms single pulse, 10 ms single pulse, dual 5 ms pulse, and reciprocal pulse (absolute …