Hydrodynamic Focused Passive Separation Under Continuous Flow In A Microfluidic Chip, Jad Kanbar
Hydrodynamic Focused Passive Separation Under Continuous Flow In A Microfluidic Chip, Jad Kanbar
Master's Theses
A continuous flow, passive separation device was designed using an equivalent circuit to create variable flow rates for hydrodynamic focusing to drain channels and collection outlets. By varying the diameter of the sample inlet connection into the reservoir, the particle position was influenced significantly, which enabled desired separations. Additionally it was noted that the relative, horizontal position of the inlet also had a significant influence on particle position within the device. A dimensionless number, the Characteristic Sample Inlet, was developed to relate geometric properties of the inlet reservoir to downstream particle distribution. It was found that a 2:1 ratio between …