Artificial --Materials --Testing.

Articles 1 - 1 of 1

Full-Text Articles in Biomedical Engineering and Bioengineering

Development Of A Gel Content Analysis Protocol For Ultra High Molecular Weight Polyethylene, Max Harold Jeanty Oct 1996

Development Of A Gel Content Analysis Protocol For Ultra High Molecular Weight Polyethylene, Max Harold Jeanty

Theses

A modified procedure based upon ASTM (D 2765 - 90) to determine the gel content of crosslinked polyethylene plastics was developed. The new procedure was applied to ram extruded Ultra High Molecular Weight (UHMWPE) implant components processed under different manufacturing conditions. This new evaluation tool provides consistent results with greater accuracy and consistency compared to the American Society for the Testing of Materials (ASTM) method. Using the same tool, several relationships between the UHMWPE processing conditions and gel content were established. The amount of gel content is related to the crosslinked material. It has been shown that the more crosslinking …