Open Access. Powered by Scholars. Published by Universities.®

Biomedical Engineering and Bioengineering Commons

Open Access. Powered by Scholars. Published by Universities.®

Western University

2011

Series

Medical Specialties

Articles 1 - 1 of 1

Full-Text Articles in Biomedical Engineering and Bioengineering

A Comparison Of Two Headless Compression Screws For Operative Treatment Of Scaphoid Fractures, Ruby Grewal, Joseph Assini, David Sauder, Louis Ferreira, Jim Johnson, Kenneth Faber Jun 2011

A Comparison Of Two Headless Compression Screws For Operative Treatment Of Scaphoid Fractures, Ruby Grewal, Joseph Assini, David Sauder, Louis Ferreira, Jim Johnson, Kenneth Faber

Surgery Publications

PURPOSE: The purpose of this study was to compare the interfragmentary compression force across a simulated scaphoid fracture by two commonly used compression screw systems; the Acutrak 2 Standard and the 3.0 mm Synthes headless compression screw.

METHODS: Sixteen (8 pairs; 6 female, 2 male) cadaver scaphoids were randomly assigned to receive either the Acutrak 2 or Synthes screw with the contralateral scaphoid designated to receive the opposite. Guide wires were inserted under fluoroscopic control. Following transverse osteotomy, the distal and proximal fragments were placed on either side of a custom load cell, to measure interfragmentary compression. Screws were placed …