High Power, Low Frequency Ultrasound: Meniscal Tissue Interaction And Ablation Characteristics, Brendan O'Daly, Edmund Morris, Graham Gavin, Conor Keane, John O Byrne, Garrett Mcguinness Jan 2011

High Power, Low Frequency Ultrasound: Meniscal Tissue Interaction And Ablation Characteristics, Brendan O'Daly, Edmund Morris, Graham Gavin, Conor Keane, John O Byrne, Garrett Mcguinness

Articles

Abstract—This study evaluates high power low frequency ultrasound transmitted via a flat vibrating probe tip as an alternative technology for meniscal debridement in the bovine knee. An experimental force controlled testing rig was constructed using a 20 kHz ultrasonic probe suspended vertically from a load cell. Effect of variation in amplitude of distal tip displacement (242–494 mm peak-peak) settings and force (2.5–4.5 N) on tissue removal rate (TRR) and penetration rate (PR) for 52 bovine meniscus samples was analyzed. Temperature elevation in residual meniscus was measured by embedded thermocouples and histologic analysis. As amplitude or force increases, there is a …