In Vitro Validation Of Finite-Element Model Of Aaa Hemodynamics Incorporating Realistic Outlet Boundary Conditions, Ethan Kung, Andrea S. Les, Francisco Medina, Ryan B. Wicker, Michael V. Mcconnell, Charles A. Taylor Feb 2011

In Vitro Validation Of Finite-Element Model Of Aaa Hemodynamics Incorporating Realistic Outlet Boundary Conditions, Ethan Kung, Andrea S. Les, Francisco Medina, Ryan B. Wicker, Michael V. Mcconnell, Charles A. Taylor

Publications

The purpose of this study is to validate numerical simulations of flow and pressure in an abdominal aortic aneurysm (AAA) using phase-contrast magnetic resonance imaging (PCMRI) and an in vitro phantom under physiological flow and pressure conditions. We constructed a two-outlet physical flow phantom based on patient imaging data of an AAA and developed a physical Windkessel model to use as outlet boundary conditions. We then acquired PCMRI data in the phantom while it operated under conditions mimicking a resting and a light exercise physiological state. Next, we performed in silico numerical simulations and compared experimentally measured velocities, flows, and …