A Database For Estimating Organ Dose For Chest And Head Ct Scans For Arbitrary Spectra And Angular Tube Current Modulation, Franco Rupcich, Andreu Badal, Iacovos Kyprianou, Taly Gilat Schmidt
A Database For Estimating Organ Dose For Chest And Head Ct Scans For Arbitrary Spectra And Angular Tube Current Modulation, Franco Rupcich, Andreu Badal, Iacovos Kyprianou, Taly Gilat Schmidt
Biomedical Engineering Faculty Research and Publications
Purpose: The purpose of this study was to develop a database for estimating organ dose in a voxelized patient model for coronary angiography and brain perfusion CT acquisitions with any spectra and angular tube current modulation setting. The database enables organ dose estimation for existing and novel acquisition techniques without requiring Monte Carlo simulations. Methods: The study simulated transport of monoenergetic photons between 5 and 150 keV for 1000 projections over 360◦ through anthropomorphic voxelized female chest and head (0◦ and 30◦ tilt) phantoms and standard head and body CTDI dosimetry cylinders. The simulations resulted in tables of normalized dose …