Open Access. Powered by Scholars. Published by Universities.®

Biomedical Engineering and Bioengineering Commons

Open Access. Powered by Scholars. Published by Universities.®

Operations Research, Systems Engineering and Industrial Engineering

2016

Hemi-Toe

Articles 1 - 1 of 1

Full-Text Articles in Biomedical Engineering and Bioengineering

Analysis Of A Clinically Failed, Mechanically Intact, Hemi-Toe Implant, Swetha Varadharajan, Richard T. Laughlin, Tarun Goswami Jan 2016

Analysis Of A Clinically Failed, Mechanically Intact, Hemi-Toe Implant, Swetha Varadharajan, Richard T. Laughlin, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

In this case study a clinically failed, mechanically intact, hemi-toe device was investigated. The clinical indication of the failure constitutes radiolucent line indicating loosening of the implant and possible interactions among the bone-implant initiated by osteolysis may become a factor, producing pain, inflammatory reactions, deformity and discomfort. The patient file was not available to determine these parameters. The device was titanium coated, Cobalt Chromium alloy used for making hemi-toe. Degeneration of hemi implant was due to spalling of the coating from the surface, causing loosening of stem from bone. However, pre-removal X-ray films were not available to confirm. Mechanism by …