Electroporation Safety Factor Of 300 Nanosecond And 10 Millisecond Defibrillation In Langendorff-Perfused Rabbit Hearts, Johanna U. Neuber, Andrei G. Pakhomov, Christian W. Zemlin
Electroporation Safety Factor Of 300 Nanosecond And 10 Millisecond Defibrillation In Langendorff-Perfused Rabbit Hearts, Johanna U. Neuber, Andrei G. Pakhomov, Christian W. Zemlin
Bioelectrics Publications
Aims
Recently, a new defibrillation modality using nanosecond pulses was shown to be effective at much lower energies than conventional 10 millisecond monophasic shocks in ex vivo experiments. Here we compare the safety factors of 300 nanosecond and 10 millisecond shocks to assess the safety of nanosecond defibrillation.
Methods and results
The safety factor, i.e. the ratio of median effective doses (ED50) for electroporative damage and defibrillation, was assessed for nanosecond and conventional (millisecond) defibrillation shocks in Langendorff-perfused New Zealand white rabbit hearts. In order to allow for multiple shock applications in a single heart, a pair of needle electrodes …