Shelf-Life Of Bioprosthetic Heart Valves: A Structural And Mechanical Study, Maryse Julien, Dany Letouneau, Yves Marois, Alain Cardou, Martin King, Robert Guidoin, Debbie Chachra, J. Lee
Shelf-Life Of Bioprosthetic Heart Valves: A Structural And Mechanical Study, Maryse Julien, Dany Letouneau, Yves Marois, Alain Cardou, Martin King, Robert Guidoin, Debbie Chachra, J. Lee
Debbie Chachra
This study was undertaken to evaluate the influence of storage conditions on the shelf-life of porcine bioprosthetic valves. Fifty-five unimplanted porcine bioprostheses have been evaluated. The valves were stored in 0.5% buffered glutaraldehyde solution for different periods of time (7, 23 and 32 months). Twenty-eight valves were refrigerated while the remaining valves were stored at room temperature. The pH of the glutaraldehyde solution at room temperature decreased with time of storage, while that kept in the refrigerator remained stable over the course of the study. Macroscopic observations showed that the valve tissues kept at room temperature, especially for the periods …