Focal Osteolysis At The Junctions Of A Modular Stainless-Steel Femoral Intramedullary Nail, Darron M. Jones, Lawrence Marsh, James V. Nepola, Joshua J. Jacobs, Anastasia K. Skipor, Robert M. Urban, Jeremy L. Gilbert, Josheph A. Buckwater Jan 2001

Focal Osteolysis At The Junctions Of A Modular Stainless-Steel Femoral Intramedullary Nail, Darron M. Jones, Lawrence Marsh, James V. Nepola, Joshua J. Jacobs, Anastasia K. Skipor, Robert M. Urban, Jeremy L. Gilbert, Josheph A. Buckwater

Biomedical and Chemical Engineering - All Scholarship

Background:

During routine follow-up of patients treated with a three-piece stainless-steel modular femoral nail, osteolysis and periosteal reaction around the modular junctions of some of the nails were noted on radiographs. The purpose of this study was to evaluate the prevalence, etiology, and clinical relevance of these radiographic findings.

Methods:

Forty-four femoral fractures or nonunions in forty-two patients were treated with a modular stainless-steel femoral intramedullary nail. Seventeen nails were excluded, leaving twenty-seven intramedullary nails in twenty-seven patients for this study. All patients had had a femoral diaphyseal fracture; nineteen had had an acute fracture and eight, a nonunion. These …