Quantifying Cross-Scatter Contamination In Biplane Fluoroscopy Motion Analysis Systems, Janelle A. Cross, Ben Mchenry, Taly Gilat-Schmidt Oct 2015

Quantifying Cross-Scatter Contamination In Biplane Fluoroscopy Motion Analysis Systems, Janelle A. Cross, Ben Mchenry, Taly Gilat-Schmidt

Biomedical Engineering Faculty Research and Publications

Biplane fluoroscopy is used for dynamic in vivo three-dimensional motion analysis of various joints of the body. Cross-scatter between the two fluoroscopy systems may limit tracking accuracy. This study measured the magnitude and effects of cross-scatter in biplane fluoroscopic images. Four cylindrical phantoms of 4-, 6-, 8-, and 10-in. diameter were imaged at varying kVp levels to determine the cross-scatter fraction and contrast-to-noise ratio (CNR). Monte Carlo simulations quantified the effect of the gantry angle on the cross-scatter fraction. A cadaver foot with implanted beads was also imaged. The effect of cross-scatter on marker-based tracking accuracy was investigated. Results demonstrated …