I-125 Ropes Eye Plaque Dosimetry: Validation Of A Commercial 3d Ophthalmic Brachytherapy Treatment Planning System And Independent Dose Calculation Software With Gafchromic Ebt3 Films, Joel Poder, Stephanie Corde Apr 2014

I-125 Ropes Eye Plaque Dosimetry: Validation Of A Commercial 3d Ophthalmic Brachytherapy Treatment Planning System And Independent Dose Calculation Software With Gafchromic Ebt3 Films, Joel Poder, Stephanie Corde

Stéphanie Corde

Purpose: The purpose of this study was to measure the dose distributions for different Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaques loaded with I-125 (model 6711) seeds using GafChromic R EBT3 films, in order to verify the dose distributions in the Plaque SimulatorTM (PS) ophthalmic 3D treatment planning system. The brachytherapy module of RADCALC R was used to independently check the dose distributions calculated by PS. Correction factors were derived from the measured data to be used in PS to account for the effect of the stainless steel ROPES plaque backing on the 3D dose distribution. …