The Evaluation Of A 2d Diode Array In "Magic Phantom" For Use In High Dose Rate Brachytherapy Pretreatment Quality Assurance, Anthony A. Espinoza, Marco Petasecca, Iolanda Fuduli, Andrew Howie, Joseph A. Bucci, Stephanie Corde, Michael Jackson, Michael L. F Lerch, Anatoly B. Rosenfeld
The Evaluation Of A 2d Diode Array In "Magic Phantom" For Use In High Dose Rate Brachytherapy Pretreatment Quality Assurance, Anthony A. Espinoza, Marco Petasecca, Iolanda Fuduli, Andrew Howie, Joseph A. Bucci, Stephanie Corde, Michael Jackson, Michael L. F Lerch, Anatoly B. Rosenfeld
Stéphanie Corde
Purpose: High dose rate (HDR) brachytherapy is a treatment method that is used increasingly worldwide. The development of a sound quality assurance program for the verification of treatment deliveries can be challenging due to the high source activity utilized and the need for precise measurements of dwell positions and times. This paper describes the application of a novel phantom, based on a 2D 11 x 11 diode array detection system, named "magic phantom" (MPh), to accurately measure plan dwell positions and times, compare them directly to the treatment plan, determine errors in treatment delivery, and calculate absorbed dose. Methods: The …