Evaluation

2015

Articles 1 - 3 of 3

Full-Text Articles in Social and Behavioral Sciences

Evaluation Of A Partial Ring Design For The Insert Spect/Mri System, Kell Erlandsson, Debora Slavado, Alexandre B. Bousse, Brian F. Hutton Jan 2015

Evaluation Of A Partial Ring Design For The Insert Spect/Mri System, Kell Erlandsson, Debora Slavado, Alexandre B. Bousse, Brian F. Hutton

Faculty of Engineering and Information Sciences - Papers: Part A

The aim of the INSERT project is to develop a SPECT insert for a commercial MRI system, for performing simultaneous SPECT/MRI brain studies in humans. We have previously investigated various design options for the detector system, based on a complete ring of detectors. We are now considering a partial ring, due to space limitations. We have investigated the degradation in image quality with a partial ring as compared to a full ring, and the possibility of addressing the limitations by utilising MRI data during reconstruction. Noise-free data were generated by forward-projecting a cylindrical phantom with spherical inserts for a full-ring …


Comparative Evaluation Of Modern Dosimetry Techniques Near Low- And High-Density Heterogeneities, Eyad A. Alhakeem, Sami Alshaikh, Anatoly B. Rosenfeld, Sergei F. Zavgorodni Jan 2015

Comparative Evaluation Of Modern Dosimetry Techniques Near Low- And High-Density Heterogeneities, Eyad A. Alhakeem, Sami Alshaikh, Anatoly B. Rosenfeld, Sergei F. Zavgorodni

Faculty of Engineering and Information Sciences - Papers: Part A

The purpose of this study is to compare performance of several dosimetric methods in heterogeneous phantoms irradiated by 6 and 18 MV beams. Monte Carlo (MC) calculations were used, along with two versions of Acuros XB, anisotropic analytical algorithm (AAA), EBT2 film, and MOSkin dosimeters. Percent depth doses (PDD) were calculated and measured in three heterogeneous phantoms. The first two phantoms were a 30 x 30 x 30 cm3 solid-water slab that had an air-gap of 20x 2.5 x 2.35 cm3. The third phantom consisted of 30 x 30 x 5 cm3 solid water slabs, two 30 x 30 x …


Evaluation Of 3d Gamma Index Calculation Implemented In Two Commercial Dosimetry Systems, Aitang Xing, Sankar Arumugam, Shrikant Deshpande, Armia George, Philip J. Vial, Lois C. Holloway, Gary Goozee Jan 2015

Evaluation Of 3d Gamma Index Calculation Implemented In Two Commercial Dosimetry Systems, Aitang Xing, Sankar Arumugam, Shrikant Deshpande, Armia George, Philip J. Vial, Lois C. Holloway, Gary Goozee

Faculty of Engineering and Information Sciences - Papers: Part A

3D Gamma index is one of the metrics which have been widely used for clinical routine patient specific quality assurance for IMRT, Tomotherapy and VMAT. The algorithms for calculating the 3D Gamma index using global and local methods implemented in two software tools: PTW-VeriSoft (R) r as a part of OCTIVIUS 4D dosimeter systems and 3DVH (TM) from Sun Nuclear were assessed. The Gamma index calculated by the two systems was compared with manual calculated for one data set. The Gamma pass rate calculated by the two systems was compared using 3%/3mm, 2%/2mm, 3%/2mm and 2%/3mm for two additional data …