In-Phantom Film Measurements Of Two Treatment Planning Systems For Single-Fraction Spine Sbrt, Michael J. Taylor May 2022

In-Phantom Film Measurements Of Two Treatment Planning Systems For Single-Fraction Spine Sbrt, Michael J. Taylor

LSU Master's Theses

Purpose: Treatment planning accuracy for spine stereotactic body radiation therapy (SBRT) varies depending on the dose calculation algorithm utilized in the treatment planning system (TPS). This project compared the end-to-end accuracy between spine SBRT plans calculated in a convolution-superposition based TPS (TPSCS) and Monte Carlo based TPS (TPSMC) with radiochromic film measurements. The hypothesis was that TPSMC would calculate the dose gradient in the critical region between the vertebral body and the spinal cord more accurately than TPSCS.