Full-Text Articles in Health and Medical Physics

On The Acceptance, Commissioning, And Quality Assurance Of Electron Flash Units, Allison Palmiero, Kevin Liu, Julie Colnot, Nitish Chopra, Denae Neill, Luke Connell, Brett Velasquez, Albert C Koong, Steven H Lin, Peter Balter, Ramesh Tailor, Charlotte Robert, Jean-François Germond, Patrik Gonçalves Jorge, Reiner Geyer, Sam Beddar, Raphael Moeckli, Emil Schüler Feb 2025

On The Acceptance, Commissioning, And Quality Assurance Of Electron Flash Units, Allison Palmiero, Kevin Liu, Julie Colnot, Nitish Chopra, Denae Neill, Luke Connell, Brett Velasquez, Albert C Koong, Steven H Lin, Peter Balter, Ramesh Tailor, Charlotte Robert, Jean-François Germond, Patrik Gonçalves Jorge, Reiner Geyer, Sam Beddar, Raphael Moeckli, Emil Schüler

Faculty, Staff and Student Publications

Background and purpose: FLASH or ultra-high dose rate (UHDR) radiation therapy (RT) has gained attention in recent years for its ability to spare normal tissues relative to conventional dose rate (CDR) RT in various preclinical trials. However, clinical implementation of this promising treatment option has been limited because of the lack of availability of accelerators capable of delivering UHDR RT. Commercial options are finally reaching the market that produce electron beams with average dose rates of up to 1000 Gy/s. We established a framework for the acceptance, commissioning, and periodic quality assurance (QA) of electron FLASH units and present an …


A Radiotherapy Community Data-Driven Approach To Determine Which Complexity Metrics Best Predict The Impact Of Atypical Tps Beam Modeling On Clinical Dose Calculation Accuracy, Fre'etta Mae Dayo Brooks, Mallory Carson Glenn, Victor Hernandez, Jordi Saez, Hunter Mehrens, Julianne Marie Pollard-Larkin, Rebecca Maureen Howell, Christine Burns Peterson, Christopher Lee Nelson, Catharine Helen Clark, Stephen Frasier Kry May 2024

A Radiotherapy Community Data-Driven Approach To Determine Which Complexity Metrics Best Predict The Impact Of Atypical Tps Beam Modeling On Clinical Dose Calculation Accuracy, Fre'etta Mae Dayo Brooks, Mallory Carson Glenn, Victor Hernandez, Jordi Saez, Hunter Mehrens, Julianne Marie Pollard-Larkin, Rebecca Maureen Howell, Christine Burns Peterson, Christopher Lee Nelson, Catharine Helen Clark, Stephen Frasier Kry

Faculty, Staff and Student Publications

PURPOSE: To quantify the impact of treatment planning system beam model parameters, based on the actual spread in radiotherapy community data, on clinical treatment plans and determine which complexity metrics best describe the impact beam modeling errors have on dose accuracy.

METHODS: Ten beam modeling parameters for a Varian accelerator were modified in RayStation to match radiotherapy community data at the 2.5, 25, 50, 75, and 97.5 percentile levels. These modifications were evaluated on 25 patient cases, including prostate, non-small cell lung, H&N, brain, and mesothelioma, generating 1,000 plan perturbations. Differences in the mean planned dose to clinical target volumes …


Outcomes Of Breakthrough Covid-19 Infections In Patients With Hematologic Malignancies, Kelly S Chien, Christine B Peterson, Elliana Young, Dai Chihara, Elizabet E Manasanch, Jeremy L Ramdial, Philip A Thompson Oct 2023

Outcomes Of Breakthrough Covid-19 Infections In Patients With Hematologic Malignancies, Kelly S Chien, Christine B Peterson, Elliana Young, Dai Chihara, Elizabet E Manasanch, Jeremy L Ramdial, Philip A Thompson

Faculty, Staff and Student Publications

Patients with hematologic malignancies have both an increased risk for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infections and higher morbidity/mortality. They have lower seroconversion rates after vaccination, potentially leading to inferior coronavirus disease 2019 (COVID-19) outcomes, despite vaccination. We consequently evaluated the clinical outcomes of COVID-19 infections in 243 vaccinated and 175 unvaccinated patients with hematologic malignancies. Hospitalization rates were lower in the vaccinated group when compared with the unvaccinated group (31.3% vs 52.6%). However, the rates of COVID-19-associated death were similar at 7.0% and 8.6% in vaccinated and unvaccinated patients, respectively. By univariate logistic regression, females, older patients, …


Analysis Of Performance And Failure Modes Of The Iroc Proton Liver Phantom, Hunter Mehrens, Paige Taylor, Paola Alvarez, Stephen Kry Jul 2023

Analysis Of Performance And Failure Modes Of The Iroc Proton Liver Phantom, Hunter Mehrens, Paige Taylor, Paola Alvarez, Stephen Kry

Faculty, Staff and Student Publications

Purpose: To analyze trends in institutional performance and failure modes for the Imaging and Radiation Oncology Core’s (IROC’s) proton liver phantom.

Materials and Methods: Results of 66 phantom irradiations from 28 institutions between 2015 and 2020 were retrospectively analyzed. Univariate analysis and random forest models were used to associate irradiation conditions with phantom results. Phantom results included pass/fail classification, average thermoluminescent dosimeter (TLD) ratio of both targets, and percentage of pixels passing gamma of both targets. The following categories were evaluated in terms of how they predicted these outcomes: irradiation year, treatment planning system (TPS), TPS algorithm, treatment machine, number …


Quantification Of The Role Of Lead Foil In Flattening Filter Free Beam Reference Dosimetry, Song Gao, Christopher Nelson, Congjun Wang, Vindu Kathriarachchi, Michael Choi, Rishik Saxena, Robin Kendall, Peter Balter Apr 2023

Quantification Of The Role Of Lead Foil In Flattening Filter Free Beam Reference Dosimetry, Song Gao, Christopher Nelson, Congjun Wang, Vindu Kathriarachchi, Michael Choi, Rishik Saxena, Robin Kendall, Peter Balter

Faculty, Staff and Student Publications

PURPOSE: To quantify the potential error in outputs for flattening filter free (FFF) beams associated with use of a lead foil in beam quality determination per the addendum protocol for TG-51, we examined differences in measurements of the beam quality conversion factor k

METHODS: Two FFF beams, a 6 MV FFF and a 10 MV FFF, were calibrated on eight Varian TrueBeams and two Elekta Versa HD linear accelerators (linacs) according to the TG-51 addendum protocol by using Farmer ionization chambers [TN 30013 (PTW) and SNC600c (Sun Nuclear)] with traceable absorbed dose-to-water calibrations. In determining k

RESULTS: Differences in %dd(10)x …


Dosimetric Analysis Of Mr-Linac Treatment Plans For Salvage Spine Sbrt Re-Irradiation, Eun Young Han, Debra N Yeboa, Tina M Briere, Jinzhong Yang, He Wang Oct 2022

Dosimetric Analysis Of Mr-Linac Treatment Plans For Salvage Spine Sbrt Re-Irradiation, Eun Young Han, Debra N Yeboa, Tina M Briere, Jinzhong Yang, He Wang

Faculty, Staff and Student Publications

PURPOSE: We investigated the feasibility of thoracic spine stereotactic body radiotherapy (SBRT) using the Elekta Unity magnetic resonance-guided linear accelerator (MRL) in patients who received prior radiotherapy. We hypothesized that Monaco treatment plans can improve the gross tumor volume minimum dose (GTVmin) with spinal cord preservation and maintain consistent plan quality during daily adaptation.

METHODS: Pinnacle clinical plans for 10 patients who underwent thoracic spine SBRT (after prior radiotherapy) were regenerated in the Monaco treatment planning system for the Elekta Unity MRL using 9 and 13 intensity-modulated radiotherapy (IMRT) beams. Monaco adapt-to-position (ATP) and adapt-to-shape (ATS) workflow plans were generated …


Adding Customized Electron Energy Beams To Truebeam Linear Accelerators, Song Gao, Manickam Muruganandham, Weiliang Du, Jared Ohrt, Rajat J Kudchadker, Peter A Balter Jul 2022

Adding Customized Electron Energy Beams To Truebeam Linear Accelerators, Song Gao, Manickam Muruganandham, Weiliang Du, Jared Ohrt, Rajat J Kudchadker, Peter A Balter

Faculty, Staff and Student Publications

Purpose: To better meet clinical needs and facilitate optimal treatment planning, we added two new electron energy beams (7 and 11 MeV) to two Varian TrueBeam linacs.

Methods: We worked with the vendor to create two additional customized electron energies without hardware modifications. For each beam, we set the bending magnet current and then optimized other beam-specific parameters to achieve depths of 50% ionization (I50 ) of 2.9 cm for 7 MeV and 4.2 cm for the 11 MeV beam with the 15 × 15 cm2 cone at 100 cm source-to-surface distance (SSD) by using an ionization chamber profiler (ICP) …


Brain Stereotactic Radiosurgery Using Mr-Guided Online Adaptive Planning For Daily Setup Variation: An End-To-End Test, Eun Young Han, He Wang, Tina Marie Briere, Debra Nana Yeboa, Themistoklis Boursianis, Georgios Kalaitzakis, Evangelos Pappas, Pamela Castillo, Jinzhong Yang Mar 2022

Brain Stereotactic Radiosurgery Using Mr-Guided Online Adaptive Planning For Daily Setup Variation: An End-To-End Test, Eun Young Han, He Wang, Tina Marie Briere, Debra Nana Yeboa, Themistoklis Boursianis, Georgios Kalaitzakis, Evangelos Pappas, Pamela Castillo, Jinzhong Yang

Faculty, Staff and Student Publications

Online magnetic resonance (MR)-guided radiotherapy is expected to benefit brain stereotactic radiosurgery (SRS) due to superior soft tissue contrast and capability of daily adaptive planning. The purpose of this study was to investigate daily adaptive plan quality with setup variations and to perform an end-to-end test for brain SRS with multiple metastases treated with a 1.5-Tesla MR-Linac (MRL). The RTsafe PseudoPatient Prime brain phantom was used with a delineation insert that includes two predefined structures mimicking gadolinium contrast-enhanced brain lesions. Daily adaptive plans were generated using six preset and six random setup variations. Two adaptive plans per daily MR image …