Safer Ambu-Bag Manual Ventilation Through Addition Of A Second High Efficiency Particulate Air (Hepa) Filter, Robert Ries, Morgan Hutchinson, Md, Robert Pugliese, Pharmd, Bon Ku, Md, Mpp, Matt Fields, Md
Safer Ambu-Bag Manual Ventilation Through Addition Of A Second High Efficiency Particulate Air (Hepa) Filter, Robert Ries, Morgan Hutchinson, Md, Robert Pugliese, Pharmd, Bon Ku, Md, Mpp, Matt Fields, Md
House Staff Quality Improvement and Patient Safety Conference (2020-)
- Experts from the Thomas Jefferson University Health Design Lab met with Emergency Medicine and Anesthesia physicians and local engineers
- A design was envisioned in which 3-D printed connectors could be created to add an additional HEPA Filter as shown in Figure 1
- STL Files for the adapters were created based on standard HEPA filter diameters and 3-D printed
- The second HEPA filter was easily assembled by novice users using the adapters
- Ventilation adequacy using the double HEPA filter ambu-bag was similar to single HEPA filtered ambu-bag ventilation based on observed inflation of a test lung