Dual Optimization Method Of Radiofrequency And Quasistatic Field Simulations For Reduction Of Eddy Currents Generated On 7t Radiofrequency Coil Shielding., Yujuan Zhao, Tiejun Zhao, Shailesh B Raval, Narayanan Krishnamurthy, Hai Zheng, Chad T Harris, William B Handler, Blaine A Chronik, Tamer S Ibrahim Nov 2015

Dual Optimization Method Of Radiofrequency And Quasistatic Field Simulations For Reduction Of Eddy Currents Generated On 7t Radiofrequency Coil Shielding., Yujuan Zhao, Tiejun Zhao, Shailesh B Raval, Narayanan Krishnamurthy, Hai Zheng, Chad T Harris, William B Handler, Blaine A Chronik, Tamer S Ibrahim

Brain and Mind Institute Researchers' Publications

PURPOSE: To optimize the design of radiofrequency (RF) shielding of transmit coils at 7T and reduce eddy currents generated on the RF shielding when imaging with rapid gradient waveforms.

METHODS: One set of a four-element, 2 × 2 Tic-Tac-Toe head coil structure was selected and constructed to study eddy currents on the RF coil shielding. The generated eddy currents were quantitatively studied in the time and frequency domains. The RF characteristics were studied using the finite difference time domain method. Five different kinds of RF shielding were tested on a 7T MRI scanner with phantoms and in vivo human subjects. …