Design Of A Simple Device For Accurate Measurement Of Human Blood Viscosity In Oxygenated And Deoxygenated Conditions, Catherine E. Oliver, Jessica Hockla, Divya Kamireddi May 2015

Design Of A Simple Device For Accurate Measurement Of Human Blood Viscosity In Oxygenated And Deoxygenated Conditions, Catherine E. Oliver, Jessica Hockla, Divya Kamireddi

Honors Scholar Theses

The purpose of this research is to design, fabricate, and test a simple device that can accurately measure the viscosity of whole blood in both an oxygenated and a deoxygenated environment. The ideal device is easy to operate, inexpensive to fabricate, and is usable outside of a laboratory setting. The microfluidic rheometer presented here was fabricated using a wet chemical etching method. Using the channel dimensions, the known viscosity of a reference fluid, and the velocity of fluid flow of the sample and a reference fluid through the microchannels the unknown viscosity of a sample fluid is calculated.