Identifying Treatment Planning System Errors In Iroc-Houston Head And Neck Phantom Irradiations, James Kerns Aug 2016

Identifying Treatment Planning System Errors In Iroc-Houston Head And Neck Phantom Irradiations, James Kerns

Dissertations & Theses (Open Access)

Treatment Planning System (TPS) errors can affect large numbers of cancer patients receiving radiation therapy. Using an independent recalculation system, the Imaging and Radiation Oncology Core-Houston (IROC-H) can identify institutions that have not sufficiently modelled their linear accelerators in their TPS model. Linear accelerator point measurement data from IROC-H’s site visits was aggregated and analyzed from over 30 linear accelerator models. Dosimetrically similar models were combined to create “classes”. The class data was used to construct customized beam models in an independent treatment dose verification system (TVS). Approximately 200 head and neck phantom plans from 2012 to 2015 were recalculated …