Investigation Of Stereotactic Body Radiation Therapy Delivery Accuracy On An Elekta Linear Accelerator, Addie Barron Jul 2018

Investigation Of Stereotactic Body Radiation Therapy Delivery Accuracy On An Elekta Linear Accelerator, Addie Barron

LSU Master's Theses

Purpose: This work investigated the delivery accuracy of high-dose lung and spine stereotactic treatments delivered with the Elekta Infinity and Versa HD platforms. The accuracy of these platforms will be used for consideration in implementing a spine stereotactic radiosurgery (SSRS) program at Mary Bird Perkins Cancer Center.

Methods: A geometric phantom was used to perform Winston-Lutz type tests that assessed the relevant degrees of freedom (gantry, collimator, and couch) of the delivery system. A lung stereotactic body radiation therapy (SBRT) and spine SRS treatment plan were generated for use in end-to-end testing. Delivery accuracy was tested using a novel diode …


Optimization Of Decision Making In Personalized Radiation Therapy Using Deformable Image Registration, Jason Vickress Jul 2018

Optimization Of Decision Making In Personalized Radiation Therapy Using Deformable Image Registration, Jason Vickress

Electronic Thesis and Dissertation Repository

Cancer has become one of the dominant diseases worldwide, especially in western countries, and radiation therapy is one of the primary treatment options for 50% of all patients diagnosed. Radiation therapy involves the radiation delivery and planning based on radiobiological models derived primarily from clinical trials. Since 2015 improvements in information technologies and data storage allowed new models to be created using the large volumes of treatment data already available and correlate the actually delivered treatment with outcomes. The goals of this thesis are to 1) construct models of patient outcomes after receiving radiation therapy using available treatment and patient …


Design & Delivery Of Automated Winston-Lutz Test For Isocentric & Off-Axis Delivery Stability Utilizing Truebeam Developer Mode & Electronic Portal Imaging Device, Mahmoud Mohammad Yaqoub May 2018

Design & Delivery Of Automated Winston-Lutz Test For Isocentric & Off-Axis Delivery Stability Utilizing Truebeam Developer Mode & Electronic Portal Imaging Device, Mahmoud Mohammad Yaqoub

UNLV Theses, Dissertations, Professional Papers, and Capstones

The uncertainties in treatment delivery cannot be ignored in radiation therapy. Thus, the quality assurance QA tests are very important task of the medical physicist in clinical practice. Assuring the coincidence between the mechanical isocenter of the Linear Accelerator (LINAC) and its radiation beams isocenter is one of the most important qualities need to be tested, and the Winston Lust (WL) test is the

most popular technique to perform this task, especially for the treatment modalities which need high precision in beam delivery such as the stereotactic radiosurgery/stereotactic body radiotherapy (SRS/SBRT). The linear accelerator-based SRS/SBRT is a well-established method in …


Development Of 3d-Printed Patient Specific Bolus For Clinical Use In Total Scalp Irradiation, Garrett Baltz May 2018

Development Of 3d-Printed Patient Specific Bolus For Clinical Use In Total Scalp Irradiation, Garrett Baltz

Dissertations & Theses (Open Access)

Total scalp irradiation (TSI) is a specialized radiation therapy technique that aims to deliver a uniform dose to the entire scalp. Original electron-based TSI techniques had limited homogeneity due to hot and cold spots created at field junctions due to the multiple matched fields that were required to treat the entire scalp. The transition to photon volumetric-modulated arc therapy based TSI techniques has improved homogeneity, to the point where non-conformal bolus is now a limiting factor. Bolus is required to build-up full dose to the scalp surface in total scalp irradiation. Creating bolus that is conformal to the scalp is …