Finding Optimal Opa-Iblc Derivatized Amino Acid Analysis Parameters With Three-Dimensional Fluorometric Scans For Hplc Method Validation, Aaron Jacobs
Honors Theses
Three-dimensional fluorometric scans of OPA-IBLC derivatized glutamic acid and histidine solutions revealed that decreasing concentration maintained maximum λem but produced different maximum λex. In pH 0.4 M sodium borate buffer (pH 10.4), 50 μM derivatized amino acid solutions had maxima around 350- to 450-nm λex-λem where 12.5 μM solutions had maxima around 230- to 450-nm λex- λem. In the lower concentration, the λex peak at 230 nm was about 80% higher than 350 nm. Thus, collecting spectra with 230 nm λex would provide better LOD and LOQ. Results inform HPLC standard concentrations since differential quenching effects may produce non-linear calibration …