Open Access. Powered by Scholars. Published by Universities.®

Engineering Commons

Open Access. Powered by Scholars. Published by Universities.®

Articles 1 - 5 of 5

Full-Text Articles in Engineering

Investigation Of Retrieved Cardiac Devices, Anmar Salih, Iosif Papadakis Ktistakis, Spyridon Manganas, Abdul Wase, Tarun Goswami Dec 2019

Investigation Of Retrieved Cardiac Devices, Anmar Salih, Iosif Papadakis Ktistakis, Spyridon Manganas, Abdul Wase, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

Damage assessment of lead and pulse generator with various exposure times is important in the development of cardiac devices. Approximately, 92.1 million patients in the US suffer from cardiovascular diseases with an estimated healthcare cost of over $300 billion and at least one million with implantable cardiac devices. These devices are complex and composed on multiple levels and present challenges while assessing the damage. However, the study on the analysis of cardiac devices may lend insight into common damage patterns and improve future cardiac devices design. The objective of this work is to perform a thorough in vivo damage assessment …


Investigation Of A Neuro-Stimulator Retrieved Posthumously, Megan Markl, Rodney J. Gutherie, Tarun Goswami Dec 2019

Investigation Of A Neuro-Stimulator Retrieved Posthumously, Megan Markl, Rodney J. Gutherie, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

A neurostimulator was investigated in this paper posthumously. Device was presented to our anatomical gift program. Investigation was multi-fold and contained visual inspection, using an optical microscope, and mechanical and electrical testing of leads and its insulator. It was concluded that the device could have been damaged during implantation, in vivo, during removal, and/or during transportation to author’s laboratories. The damage observed on the lead insulation is similar to that which can occur due to anchoring of the lead and hardening due to oxidation. Insulation stiffness was determined to be 1/10 of new insulator. The results reported here on the …


Simulation Of Skull Fracture Due To Falls, Anthony Vicini, Tarun Goswami Dec 2019

Simulation Of Skull Fracture Due To Falls, Anthony Vicini, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

This study presents novel predictive equations for von Mises stress values of bones in the frontal and lateral regions of the skull. The equations were developed based on results of a finite element model developed during this research. The model was validated for frontal and lateral loading conditions with input values mimetic to fall scenarios. Using neural network processing of the information derived from the model achieved R2 values of 0.9990 for both the stress and deflection. Based on the outcome of the fall victims, a threshold von Mises stress of 40.9 to 46.6 MPa was found to indicate skull …


Deep Brain Stimulation: Prediction Model For Volume Of Tissue Stimulated, Melissa Jones, Tarun Goswami Nov 2019

Deep Brain Stimulation: Prediction Model For Volume Of Tissue Stimulated, Melissa Jones, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

An attempt has been made to understand deep brain stimulation in humans via marketable devices that are approved by the US Food and Drug Administration. The electrode characteristics were presented to influence the volume of tissue stimulated. Experimental data were digitized and VTS data were presented a function of electrode diameter, aspect ratio, applied current, pulse duration, voltage and frequency. Amongst other trends voltage and VTS were found to exhibit a three-stage relation, where Stage II characteristics were represented by a linear equation. It is in this stage, the VTS was found to be stable and where the most effective …


Stress Shielding In Cemented Hip Implants Assessed From Computed Tomography, Bharadwaj Cheruvu, Indresh Venkatarayappa, Tarun Goswami May 2019

Stress Shielding In Cemented Hip Implants Assessed From Computed Tomography, Bharadwaj Cheruvu, Indresh Venkatarayappa, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

Background: Aseptic loosening is the major cause of revisions for hip replacement. This mode of failure is often caused by stress shielding. Stress shielding in the femur occurs when some of the loads are taken by the prosthesis and shielded from going to the bone. There is little information regarding the stress shielding among cemented hip implants. Purpose: The purpose of this study is to investigate the effect of stress shielding on the proximal femur with a femoral prosthesis. Methods: A patient had undergone open reduction and internal fixation (ORIF) due to a comminuted reversed oblique fracture of the right …