Wear Of Polyethylene And Hylamer On Cobalt-Chromium : A Knee Simulator Study, Alessandro F. Canonaco
Wear Of Polyethylene And Hylamer On Cobalt-Chromium : A Knee Simulator Study, Alessandro F. Canonaco
Theses
Two tests were conducted to examine the wear characteristics of tibial bearings used in total knee replacement systems. Each test consisted of six A/P Glide Tibial Bearings each having a conical control arm. The plastic portion of the conical bearings were all made of UHMWPe. Each of these bearing systems was mounted onto a Co-Cr alloy tibial platform and Co-Cr alloy LCS (low contact stress) femoral component . These test samples were mounted onto the New Jersey Mark III Knee Simulator System. The simulator was configured to produce flexion and axial rotation to simulates normal gait. Each test ran at …