Nitric oxide

Articles 1 - 1 of 1

Full-Text Articles in Engineering

Fabrication And In Vivo Thrombogenecity Testing Of Nitric Oxide Generating Artificial Lungs, Kagya Amoako, Patrick Montoya, Terry C. Major, Ahmed B. Suhaib, Hitesh Handa, David O. Brant, Mark E. Meyerhoff, Robert H. Bartlett, Keith E. Cook Dec 2013

Fabrication And In Vivo Thrombogenecity Testing Of Nitric Oxide Generating Artificial Lungs, Kagya Amoako, Patrick Montoya, Terry C. Major, Ahmed B. Suhaib, Hitesh Handa, David O. Brant, Mark E. Meyerhoff, Robert H. Bartlett, Keith E. Cook

Mechanical and Industrial Engineering Faculty Publications

Hollow fiber artificial lungs are increasingly being used for long-term applications. However, clot formation limits their use to 1–2 weeks. This study investigated the effect of nitric oxide generating (NOgen) hollow fibers on artificial lung thrombogenicity. Silicone hollow fibers were fabricated to incorporate 50 nm copper particles as a catalyst for NO generation from the blood. Fibers with and without (control) these particles were incorporated into artificial lungs with a 0.1 m2 surface area and inserted in circuits coated tip-to-tip with the NOgen material. Circuits (N = 5/each) were attached to rabbits in a pumpless, arterio-venous configuration and run for …