Computer Modeling And Simulation Of Implantable Medical Device Heating Due To Mri Gradient Coil Fields, Bryan David Stem Apr 2014

Computer Modeling And Simulation Of Implantable Medical Device Heating Due To Mri Gradient Coil Fields, Bryan David Stem

Open Access Theses

For patients with implantable medical devices, the ability to safely undergo MRI scanning is critical to ensuring the highest standard of care. The gradient coils of an MRI generate kilohertz frequency, time varying magnetic fields. These magnetic fields induce a voltage on the external case of metallic, implantable medical devices through electromagnetic induction. Since the magnetic field generated by a gradient coil is time varying, the induced voltage results in the flow of eddy currents which can cause heating effects. These heating effects have been successfully modeled using ANSYS Maxwell and ANSYS Mechanical software packages.

The multi-physics simulation and solution …


Effects Of Blood Flow On The Heating Of Cardiac Stents Due To Radio Frequency Fields, Nate Ian Elder Jan 2013

Effects Of Blood Flow On The Heating Of Cardiac Stents Due To Radio Frequency Fields, Nate Ian Elder

Open Access Theses

A safety concern during MRI scans with implanted medical devices is heating induced by the incident RF field. This research was performed to better understand the heating of cardiac stents during MRI. Heating of cardiac stents tends to occur at their ends. The temperature rise will be affected by blood flow through the lumen of the stent. In this work, an experiment was performed to simulate heating of a cardiac stent in the presence of blood flow during exposure to the electric field induced by a 64 MHz magnetic field, which corresponds to MRI at 1.5 T. The test procedure …