Open Access. Powered by Scholars. Published by Universities.®

Engineering Commons

Open Access. Powered by Scholars. Published by Universities.®

Biomedical Engineering and Bioengineering

Biomedical, Industrial & Human Factors Engineering Faculty Publications

Finite element

Publication Year

Articles 1 - 2 of 2

Full-Text Articles in Engineering

Biomechanical Behavior Of A Variable Angle Locked Tibiotalocalcaneal Construct, Farah Hamandi, Gerard Simon, Richard T. Laughlin, Tarun Goswami Mar 2020

Biomechanical Behavior Of A Variable Angle Locked Tibiotalocalcaneal Construct, Farah Hamandi, Gerard Simon, Richard T. Laughlin, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

This paper examines the mechanics of the tibiotalocalcaneal construct made with a PHILOS plating system. A failed device consisting of the LCP plate and cortical, locking, and cannulated screws was used to perform the analysis. Visual, microstructure, and fractographic examinations were carried out to characterize the fracture surface topology. These examinations revealed the presence of surface scratching, inclusions, discoloration, corrosion pits, beach marks, and cleavage and striations on the fracture surface. Further examination of the material crystallography and texture revealed an interaction of S, Ni, and Mo-based inclusions that may have raised pitting susceptibility of the device made with Stainless …


Failure Analysis Of Philos Plate Construct Used For Pantalar Arthrodesis Paper Ii-Screws And Fem Simulations, Farah Hamandi, Richard T. Laughlin, Tarun Goswami Apr 2018

Failure Analysis Of Philos Plate Construct Used For Pantalar Arthrodesis Paper Ii-Screws And Fem Simulations, Farah Hamandi, Richard T. Laughlin, Tarun Goswami

Biomedical, Industrial & Human Factors Engineering Faculty Publications

A fractured stainless steel 3.5 mm proximal humerus internal locking system (PHILOS) plate and screws were investigated in this paper. This plate was used for ankle arthrodesis of a 68-year-old female with a right ankle deformity. Both the plate and screws were considered in this investigation. Optical and scanning electron microscopes (SEM) were used to document fracture surface characteristics, such as extensive scratching, plastic deformation, rubbed surfaces, discoloration, and pitting, along with cleavage, secondary cracking, deposits of debris, striations, and dimples. Indications of these features show that the plate failed by corrosion fatigue, however, overloading separated the screw(s) in two …