Optimal Hemoglobin Extinction Coefficient Data Set For Near-Infrared Spectroscopy, Yue Zhao, Lina Qiu, Yunlong Sun, Chong Huang, Ting Li Nov 2017

Optimal Hemoglobin Extinction Coefficient Data Set For Near-Infrared Spectroscopy, Yue Zhao, Lina Qiu, Yunlong Sun, Chong Huang, Ting Li

Biomedical Engineering Faculty Publications

Extinction coefficient (ε) is a critical parameter for quantification of oxy-, deoxy-, and total-hemoglobin concentrations (Δ[HbO2], Δ[Hb], Δ[tHb]) from optical measurements of Near-infrared spectroscopy (NIRS). There are several different ε data sets which were frequently used in NIRS quantification. A previous study reported that even a small variation in ε could cause a significant difference in hemodynamic measurements. Apparently the selection of an optimal ε data set is important for NIRS. We conducted oxygen-state-varied and blood-concentration-varied model experiments with 57 human blood samples to mimic tissue hemodynamic variations. Seven reported ε data sets were evaluated by comparisons between …