Open Access. Powered by Scholars. Published by Universities.®

Medicine and Health Sciences Commons

Open Access. Powered by Scholars. Published by Universities.®

2010

Series

Thomas Jefferson University

Department of Anesthesiology Faculty Papers

Articles 1 - 1 of 1

Full-Text Articles in Medicine and Health Sciences

High Inspired Co2 Capnograph Reading Due To Misplaced End Tidal Co2 Sampling Port, Andrew Ng, Md, Elia Elia, Md, Lisa Luyun, Md Oct 2010

High Inspired Co2 Capnograph Reading Due To Misplaced End Tidal Co2 Sampling Port, Andrew Ng, Md, Elia Elia, Md, Lisa Luyun, Md

Department of Anesthesiology Faculty Papers

In conclusion, if abnormal capnography is seen where the baseline inspiratory CO2 is higher than usual, inspect the circuit for rebreathing and consider examining the position of ETCO2 sampling port. We recommend confirming ETCO2 sampling port is connected immediately distal to Y-piece before starting a case for accurate CO2 measurement.